UNION Instruments GmbH
INCA Analyzer Series
– Instruction Manual
Page 38 of 59
Calibration
5.1.5
Manual calibration
Manual calibration can be performed using the user interface of the display. Under
Commands
menu, there perform calibration using purge gas.
Calibration using calibration gas
The calibration process can manually be started by a command from user interface of the
display.
Commands
Cal. purge gas
Cal. gas I
Reset calibration
Under the
Commands
menu, press
▲ and ▼ to
highlight
Cal. purge gas
and press
MENU
/ .
Press
MENU
/ to execute the command.
Cal. purge gas
Press ENTER...
Executing...
OK
Cal. purge gas 7:14
Channel :
1/2
CH
4
: 0.03 Vol.%
CO
2
: 0.00
Vol.%
Commands
Cal. purge gas
Cal. gas I
Reset calibration
Under
Commands
menu, press
▲ and ▼ to highlight
Cal. gas I
and press
MENU
/ .
Press
MENU
/ to execute the command.
Cal. gas I
Press ENTER...
Executing...
OK
Cal. gas I
9:47
Channel :
1/2
Fr. 30.10.2009
12:10:12