UNION Instruments GmbH
INCA Analyzer Series
– Instruction Manual
Page 24 of 59
System main menu
4.2
Analyzer display during warm-up
Figure 4-5: Start-up display
The analyzer with NDIR (infra red) sensor module or a Parox sensor needs warm-up time
(usually 10-20 minutes) to meet their operating temperatures. The figure above shows the
temperature of NDIR sensor group and the status of the Parox sensor. Since the T(IR)
already meets its operating temperature, “OK” is shown in the display.
Operating temperature:
NDIR sensor module
: 49 or 64 °C
If the Parox sensor reaches its operating temperature, the display shows T(POX) = 0x0000.
Analyzers without those parts do not need any warm-up time and therefore can be operated
immediately. The device starts the actual operating phase after completing the warm-up
process. These phases run through and start again from the beginning in a cycle.
WARMUP
*
T(IR) : 49.2°C - OK
---
T(POX) : 0x0400 -