UNION Instruments GmbH
INCA Analyzer Series
– Instruction Manual
Page 56 of 59
Appendix
0x030A
SENS CAL STATUS SPAN EC1
Span calibration could not be performed on EC channel 1
--- currently not used ---
0x030B
SENS CAL STATUS ZERO EC2
Zero calibration could not be performed on EC channel 2
--- currently not used ---
0x030C
SENS CAL STATUS SPAN EC2
Span calibration could not be performed on EC channel 2
--- currently not used ---
0x030D
SENS EC PRESSURE AIR
Min. pump pressure not reached – check pumps
0x030E
SENS EC PRESSURE GAS
Min. pump pressure not reached – check pumps
0x030F
SENS EC3 PRESSURE AIR
Min. pump pressure not reached – check pumps
0x0310
SENS EC3 PRESSURE GAS
Min. pump pressure not reached – check pumps
0x0311
SENS EC4 PRESSURE AIR
Min. pump pressure not reached – check pumps
0x0312
SENS EC4 PRESSURE GAS
Min. pump pressure not reached – check pumps
0x0313
SENS CALC. VALUE
NOT DEFINED
Firmware development error
Method not defined in child class
0x0314
SENS FATAL CHNL.
ERR NOT DEF.
Firmware development error
Method not defined in child class
0x0315
SENS ABORT MEAS.
NOT DEF.
Firmware development error
Method not defined in child class
0x0316
SENS FACTORY
CALIB. NOT DEF.
Firmware development error
Method not defined in child class
0x0317
SENS IR CALIB. NOT READ
Firmware development error
Sensor calibration was not read before trying to reset
calibration to factory settings
0x0318
SENS IR PRESS.
PROCESS GAS
Error for continuous measurement instrument
Minimal required pressure in process gas stream not
reached – pressure measured through absolute pressure
sensors of IR electronic
0x0319
SENS IR PRESS. CALIB GAS
Error for continuous measurement instrument
Minimal required pressure in calibration gas stream not
reached – pressure measured through absolute pressure
sensors of IR electronic
0x031A
SENS SET FACT.
CALIB NOT DEF.
Firmware development error
Method not defined in child class
0x031B
SENS TYPE NOT DEFINED
Sensor type not defined – error in calibration
Fatal sensor errors
Error code
Display description
Full description
0x0380
SENS GLOB ERROR
Error is returned by a sensor
0x0381
SENS GLOB ERROR IO
Global error occurred in IO board
0x0382
SENS GLOB ERROR IR
Global error occurred in IR board
0x0383
SENS GLOB ERROR EC
Global error occurred in EC board
0x0384
SENS CHNL. ERROR IR1
Fatal channel error occurred in IR channel 1
0x0385
SENS CHNL. ERROR IR2
Fatal channel error occurred in IR channel 2
0x0386
SENS CHNL. ERROR EC1
Fatal channel error occurred in EC channel 1
0x0387
SENS CHNL. ERROR EC2
Fatal channel error occurred in EC channel 2
0x0388
SENS GLOB ERROR EC3
Global error occurred in EC3 board
0x0389
SENS GLOB ERROR EC4
Global error occurred in EC4 board
0x038A
SENS ANLG. OUTPUT
NOT VALID
Firmware development error
Analog output number too large
0x038B
SENS EC CALIB PTR NOT SET
Firmware development error
EC span calibration aborted due to missing pointer
0x038C
SENS EC CALIB
ABORT ZERO
Zero calibration for EC sensor aborted