Figure 6-19 Display Selected Parameters
Note
: Parameter selection is to measure parameter immediately. There are 19 types of parameter
measurements for UTD1000CL and 22 types available in UTD1000DL. Under general circumstances, not all
parameters, but just some of them are required to be measured. Therefore the users can customize their
measurements as expected. The software can select up to 4 parameters displayed on the screen.
Auto Parameter Measurements
Figure 6-20 Voltage Measurement
The voltage parameters that the Oscilloscope can automatically measure include:
Maximum Value (Vmax): Voltage measured from the highest point of waveform to the zero ground
Minimum Value (Vmin): Voltage measured from the lowest point to the ground
Top Value (Vtop): Voltage measured from the flat top to the ground.
Base Value (Vbase): Voltage measured from the base of the waveform to the ground
Middle Value (Vmid): a half of amplitude value
Peak-to-Peak Value (Vpp): Voltage measured from the highest point to the lowest point of waveform
Amplitude (Vamp): Votage measured from the top to the base of the waveform
Overshoot: (Maxi-Top)/Amplitudex100%
Preshoot: (Min-Low)/Amplitudex100%