User’s Manual MSO/UPO2000 Seri es
81 / 137
Instruments.uni-trend.com
Table 13-3 Template Setting
Function Menu
Setting
Description
Reference
Waveform
CH1, CH2, CH3, CH4,
Local REF, USB REF
Specify a channel waveform in CH1
~
CH4, DSO REF, USB REF and plus the
horizontal and vertical tolerance as the conditions for creating a template
Horizontal
Tolerance
1
~
100
Adjusting Multipurpose knob to select the range of horizontal tolerance. The
default value is 5.
Vertical Tolerance 1
~
100
Adjusting Multipurpose knob to select the range of vertical tolerance. The
default value is 5.
REF Loading
Loading reference waveform as the reference waveform for creating
template.
(6)
Operating
After all above setting is completed, press operating key to run the pass/fail test.
13.4 System Update
This series can use USB to update program. It’s convenient and flexible.
USB update has two methods. Method 1: Turn on the oscilloscope to detect USB update; Method 2:
utility-update-detect USB update file and select the file to update. The specific steps refer to the
following.
Method 1
a.
Press UTILITY key to enter auxiliary function menu, press system key to check the system
information to find the model name, software and hardware version information.
b.
Download the update file from UNI-T official website or ask UNI-T distributor to provide the
update file. The update file is the same as the model and hardware version of the instrument, the
software version is higher than the version of the instrument. Save the update file in the root
directory of the USB.
c.
The instrument is in shut down state. Insert USB and boot the instrument, wait it to enter the
upgrade interface, and then press F1 to confirm to update.
d.
Updating process requires about 5min. After it finished, turn off the instrument and plug out USB.
e.
Reboot the instrument to check whether the system information is same as the provided version.
If it is the same that means the update is successful.
Method 2
a.
Press UTILITY key to enter auxiliary function menu, press system key to check the system
information to find the model name, software and hardware version information.
b.
Download the update file from UNI-T official website or ask UNI-T distributor to provide the
update file. The update file is the same as the model and hardware version of the instrument, the
software version is higher than the version of the instrument. Save the update file in the root