SERVICE PROCEDURES
1) CLEAN CONDENSER
Using a brush attachment, vacuum the condenser at the back of the unit.
2) CLEAR BLOCKED SOLVENT FLOW PATH
CAUTION:
WEAR SAFETY GOGGLES.
A blockage in the solvent fl ow path (Vapour Outlet - Condenser - Solvent Outlet Tube) can cause solvent
to leak. To determine the location of the blockage, fi rst check the Solvent Outlet Tube and the Vapour
Outlet Fitting for visible signs of blockage or damage. If the problem is not in either of these locations, the
Condenser is problably blocked. To clear the condenser, pour some clean solvent into the Vapour Outlet
and check if it comes out of the Solvent Outlet Tube. If the blockage persists, blow air at about 30 PSI (2
kg/cm2) into the Vapour Outlet. If the air comes out of the Solvent Outlet Tube, the blockage has been
cleared. If not, call for service.
3) REPLACE FUSES
Fuses are located on the Power Control Board inside the Motor Housing.
• Disconnect power supply.
• Remove the Guard Screen by unscrewing two metal screws to get access to the Motor Housing
• Unscrew the 6 screws from the Front Cover and pull it gently from the motor housing to expose the
fuses.
Note: Care should be taken not to pull the Front Cover too far as some wires may discon-
nect.
• There are two fuses for URS500 series and four fuses for URS600 series.
URS500 series:
URS600 Series:
Fuses F1: 250 V 20.0 A 314 (Fast type) for Heater
There are two each of F1 and F2 fuses.
Fuses F2: 250 V 2.0 A 312 (Slow type) for Fan Motor
• Remove the fuses from the board and, using a meter, test each one and replace as needed.
• Carefully push the Power Control Board back into the Motor Housing.
• Ensure that the wire to the computer board is secure.
• Re-install the Front Cover using all 6 screws..
• Install the Guard Screen using two metal screws.
• Close the Door and re-connect the power supply.