- 80 -
8-2 REC.
OUT
Output
This is a voltage output of leak rate during test. Connect the output signal to a strip chart
recorder to use it for recording of leak rate.
This output has two formats of 0 to 10 VDC (when linear display is set) and 0 to 8 VDC
(when log display is set) depending on the setting of the bar graph display selection mode on
the controller.
!
CAUTION
No short-circuit
Do not short the REC. OUT terminal (red) and the REC. OUT terminal
(black).
If shorted, the equipment may be damaged.
(1) Output
signal
Terminal
Contents
Signal format
REC. OUT red
Leak rate
0 to 10 VDC maximum
REC. OUT black
Leak rate output –
GND
<Output voltage of linear display>
Leak rate value
Output voltage (V)
Leak rate value
Output voltage (V)
9.9E-** 10
9.9E-04 8.0
9.0E-** 9
5.0E-04 7.5
8.0E-** 8
1.0E-04 7.1
7.0E-** 7
5.0E-05 6.5
6.0E-** 6
1.0E-05 6.1
5.0E-** 5
5.0E-06 5.5
4.0E-** 4
1.0E-06 5.1
3.0E-** 3
5.0E-07 4.5
2.0E-** 2
1.0E-07 4.1
1.0E-** 1
5.0E-08 3.5
0.0E-** 0
1.0E-08 3.1
5.0E-09
2.5
1.0E-09
2.1
5.0E-10 1.5
1.0E-10
1.1
5.0E-11
0.5
1.0E-12
0.1
* Relationship between leak rate and REC. OUT output voltage in log. display
LR = 10 (V – E)
×
10
-(11 - E)
LR : Leak
rate
V
: REC. OUT output voltage
E
: REC. OUT output voltage value from which digits
below decimal point are truncated.