Section
7
Maintenance
ULVAC CRYOGENICS INCORPORATED
7-1
Section 7
Maintenance
Scheduled maintenance ....................................................................... 7-1
Replacement of adsorber ..................................................................... 7-1
Adjusting system helium pressure ........................................................ 7-5
7.1 Scheduled maintenance
The table below shows the required maintenance on the cryopump system. Refer to the
following paragraph for details of adsorber replacement.
For maintenance of refrigerator or other components of the cryopump, see cryopump
instruction manual.
Table 7-1
Scheduled maintenance
Item
Maintenance Cycle (h)
Adsorber replacement
Max. 24000h (30000h for C30MVE
_
R)
Refrigerator maintenance
When needed
(
See cryopump instruction manual.
)
Inspection for plumbing and
wirings
12000h
NOTE:
If a special maintenance plan for your specific application is needed please consult
with our Service Engineering Division or the nearest customer support center.
NOTE:
Refer to “Disposal Considerations” when disposing used components or parts.
7.2 Replacement of adsorber
Neglecting the scheduled maintenance may shorten the product life, cause damage, and/or
destruction of the cryopump system.
CAUTION
The following replacing procedure must be undertaken by authorized personnel familiar with
the structure and risk of the product. Failure to observe this warning may result in severe
CAUTION
bodily injury or fatal accident.
Summary of Contents for C30VR Series
Page 2: ......
Page 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 16: ...S 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 18: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 108: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 110: ...Revision History R 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...