50
English
9.9 Alarm table
ALARM
DESCRIPTION
ACTION
E00 EEPROM ALARM
Firmware installed successfully.
1- Press RESET
2- Install new dataset
E01 BLACKOUT
A blackout occurred during cycle execution
and the program was stopped.
Press RESET
E02 DOOR OPEN
The door is open or unlocked.
Close the door.
Press RESET
If it persists, call the technician.
E07 NO LOCKED DOOR
The door did not lock within the set time.
Press RESET
If it persists, call the technician.
E09 NO UNLOCKED DOOR
The door did not lock within the set time.
Press RESET
If it persists, call the technician.
E11 DIRTY WATER
Check for dirt on the load and check the inlet
demineralized water.
Press RESET
If it persists, call the technician.
E20 NO COLD WATER
The cold water tap could be closed or
partially open. Check it.
Press RESET
If it persists, call the technician.
E21 NO HOT WATER
The hot water tap could be closed or
partially open. Check it.
Press RESET
If it persists, call the technician.
E22 NO DEMI WATER
The demineralized water tap could be closed
or partially open. Check it.
Press RESET
If it persists, call the technician.
E24 DRYING FAN
The drying fan might be malfunctioning,
broken or disconnected.
Press RESET
If it persists, call the technician.
E26 PRINTER ERROR
The printer is turned off, unplugged or out of
paper. Check it.
Press RESET
If it persists, call the technician.
E30 FLOWMETER 1
Chemical liquid tank 1 might be empty.
Check it.
Press RESET
If it persists, call the technician.
E31 FLOWMETER 2
Chemical liquid tank 2 might be empty.
Check it.
Press RESET
If it persists, call the technician.
E32 FLOWMETER 3
Chemical liquid tank 3 might be empty.
Check it.
Press RESET
If it persists, call the technician.
E34 PRODUCT TIMEOUT
The density of the chemical liquid might be
too high. Check it.
Press RESET
If it persists, call the technician.
E41 DRAIN TIMEOUT
The drain pipe might be blocked or clogged.
Check it.
Press RESET
If it persists, call the technician.
E43 DRY TEMPERATURE
The load might not be dried because the
minimum set temperature has not been
reached.
Press RESET
If it persists, call the technician.
E44 PREWASH TEMPERAT
A too high temperature was detected during
the prewash phase.
Press RESET
If it persists, call the technician.
E45 CHAMBER T.LIMIT
A too high temperature was detected inside
the chamber.
Press RESET
If it persists, call the technician.
E46 AIR T.LIMIT
A too high air temperature was detected.
Press RESET
If it persists, call the technician.
E47 PHASE T. LIMIT
A too high temperature was detected during
the current phase.
Press RESET
If it persists, call the technician.
E50 PROBE 1 BROKEN
Chamber probe PT1000-1 (work) might be
broken or disconnected.
Press RESET
If it persists, call the technician.