The thermodisinfector is equipped with
a product container with relative pump
(1 Fig. 15) for the detergent liquid. On
request, it can be equipped with a second
container (2/3 Fig. 15) used for the
neutralising liquid or rinse aid.
The under-the-counter thermodisinfector
is equipped with three tanks (Fig. 15), with
relative pumps, for greater capacity of
the liquids where tank 1 will contain the
detergent liquid, tank 2 the neutraliser and
tank 3 the rinse aid.
Drying
The instruments are dried internally and externally
with filtered hot air. Good drying reduces the risk of
corrosion caused by the residual humidity on the load.
Instruments with a very small internal diameter must
be dried further.
a
WARNING. The program can be interrupted
at any time. However, it must be remembered
that once interrupted, it must restart from
the beginning. Only if it is interrupted in the
drying phase can the cycle be considered
completed successfully. The instruments
must be dried.
Ö
IMPORTANT: In the EcoDrying version, the
machine will interrupt the cycle at the end
of the disinfection. It is the responsibility of
the user to dry the instruments that have
just been washed, using suitable cloths
and following the instructions contained in
the instruction manual for each individual
instrument washed. The manufacturer
declines any liability for contamination of the
instruments during the manual drying stage.
Ö
IMPORTANT: In the EcoDrying version,
the instruments must NOT be left inside
the machine, at the end of the cycle,
wet, for longer than 5 minutes. This is to
prevent the proliferation of bacteria and
microorganisms.
6.4.3 Program End
The message «Successfully completed» on the display
indicates that the program has been carried out
correctly. Unlock the door by pressing the key (5 Fig.
12) and open it.
Ö
Open the door immediately after the end of
the program to avoid condensation forming.
Check the results at the end of the cleaning process.
The instruments must be completely clean and dry.
Ö
Hollow instruments with a small internal
diameter must be dried further with air.
- Check hollow instruments and retreat them if
necessary.
- The hole (lumen) of hollow instruments must be free.
- The injectors must be correctly placed on the
connection tube in the cleaning chamber.
- The nozzles and connections to the base grid must
be firmly fixed.
If these checks are positive and the program has been
carried out without interruptions or malfunctions, the
load has been successfully cleaned and disinfected.
6.4.4 Extracting the load
At the end of the program and when extracting the
load:
Ö
Do not force to open the door to avoid
damaging the device and/or the leak of toxic
steam.
Ö
Particularly large instruments can be
very hot at the end of the program. Allow
instruments, containers and inserts to cool
down before removing them.
Failure to follow these recommendations can cause
burns.