83
Chamber clean Cleaning procedure
Important
The Chamber Clean cleaning program for class B and S for T-Edge10 autoclaves is
described below:
The chamber must be cold.
1.
All steps in this procedure must be completed without interruption.
2.
Open the autoclave chamber door.
3.
Remove instruments and any load, trays from the autoclave chamber.
4.
Place the trays and tray holder in the sink to be cleaned with a
stainless steel Safe cleaner.
5.
Place one 6g cleaning tablet (1) from the tablet packing (2) inside the
sterilization chamber at the back bottom of the cold chamber.(see
figure below). Under the 132°C program conditions. Chamber clean
program 6g Tablet position
1. Tablet
2. Tablet from packing
6.
Close the autoclave door.
7.
Fill fully (Approx. 23 liters for T-Edge 10, and 28 Liters for T-Edge 11)
clean water reservoir with
distilled water only
. (Either Automatic or
manual filling). For manual filling, select 0, Global select 1 for
automatic filling. Fill the reservoir full until the screen indicate full.
1