24
8. Check and verify that the ambient relative humidity does not exceed
80%
9. The operational altitude shall not be over 13123 ft (4000 meters).
10. Ambient pressure shall not be lower than 8.8 psi (60.5 KPa) (if the
altitude and temperature are kept in the manufacturer's instructed
ranges above, and no forced extreme negative pressure is applied near
the autoclave, then ambient pressure of 8.8 psi (60.5 KPa) or higher is
guaranteed, as it is a function of altitude and temperature).
11. Operate the autoclave only in the manner specified in the manual. If the
equipment is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired.
Connections to Utility Supplies
1. Check and verify that the power supply is a 1 phase, 110Vac ±10%,
60Hz, 10A supply.
2. Check and verify that the electrical net is protected by a current
leakage safety relay.
Storage
After the removal of the autoclave from the package, we recommend the
following:
Keep the device dry.
Keep the device away from sunlight and protect it from heat.
Initial Operation of the Device
1. Plug the power cord into the power socket.
Warning!
Mind the Power socket. Keep it and its vicinity dry. Danger
of electrocution.