59
Periodic Maintenance
Caution!
Make sure the autoclave is not hot before cleaning it
1. Replace the air filter, every 6 months or after 1000 cycles
(whichever comes first) according to 12.3.
2. Every 3 months check the door gasket for any signs of physical
damage and ask the technician to replace it if there is tear or
leakage.
9.4.1
Replacing the Air Filter
Cautions!
Before proceeding, make sure that the electric cord is
disconnected and there is no pressure in the chamber.
Use scissors to open the filter bag and not sharp blades or
pointed instrument.
Check that the new filter has not exceeded the maximum
shelf life.
Carefully un-pack the new filter and examine it for any signs
of damage.
Remove any protective packaging before inserting the filter
into place.
Carefully Insert the new filter into the housing. Do not force.
The AIR filter is located under the mounted under the cover (see below).
1. Unscrew the filter
(see below)
.