14
1.11.2
T-Edge10 fit into sterilization center cabinet - recommended dimensions:
To allow adequate cooling & ventilation of the autoclave placed inside the sterilization
center Cabinet, the following T-Edge10 Cabinet sizes are recommended:
Width - 2
2.85”- 23.65”/58-60cm; Height – 22.25”- 22.65”/56.5-57.5cm; Min. Depth 24.9”-
25.2”/63.2-64cm
1.11.3
Device Properties
Property
Dimension
External size
Width
18.9"(48cm)
Height
19"(49 cm)
Depth
22.8" (58 cm) supporting common install
base carry a 23" (60 cm) counter top
Chamber
Diameter
10"(25 cm)
Depth
18"(46 cm)
Volume
778 Ounces (23lit)
Usable chamber space
75% (~575 Ounces / ~17 L)
Max. Allowable Working Pressure (MAWP)
40.6PSI (2.8 bar)
Maximum load per item
0.45 lbs. (0.2kg)
Maximum load per tray
Unwrapped
2.67 lbs. ( 1.2kg)
Wrapped
1.6 lbs (0.72kg)
aximum Solid load
Unwrapped
13 lbs.(6kg)
Wrapped
7.7 lbs.(3.5kg)
Maximum textile load
3.3 lbs.(1.5kg)
Tray dimensions
16.6" x7.4"x0.8" (42.1cm x 18.9cm x2.05cm)
No. of trays
13 lbs (6 kg )
Net weight
110lbs (50kg)
Shipping weight
132lbs (60kg)
Floor loading requirements
165 lbs(75kg).
Mineral-free water reservoir
Max. water volume
Overflow (up to the float)
– 128 Ounces
(3.8lit)
Min. water volume
.33.8 Ounces (1lit)
The volume used by the
sterilization cycle/load having
the highest steam
consumption
Recorded 27 Ounces (800ml) were required
to sterilize full load of porous type using
"wrapped 121".
Used (waste) water
reservoir
Max. water volume
Max vol.
–./135 Ounces /(4.0 lit)
Float
– 125 Ounces (3.7lit.) max allowed for
start cycle
Safety relief valve
40 PSI (2.8 bar)
Load No. counter
Counting from 0 to 999 and nullifies.
Table 2 -Device Properties