3.11. Water Quality
Physical Characteristics and Contaminants Levels
The water supplied to the autoclave should have the physical characteristics and maximum
acceptable level of contaminants indicated in the table below:
Physical Characteristics and Maximum acceptable contaminants levels
in steam for sterilizers (According to EN 13060: 2010)
Element
Condensate – allowable content
Silicon dioxide SiO
2
≤0.1 mg/kg
Iron
≤0.1 mg/kg
Cadmium
≤0.005 mg/kg
Lead
≤ 0.05 mg/kg
All other metals except iron, cadmium, lead
≤0.1 mg/kg
Chloride (Cl)
≤0.1 mg/kg
Phosphate (P
2
O
5
)
≤0.1 mg/kg
Conductivity
≤3 μs/cm
pH value (degree of acidity)
5 to 7
Appearance
Colorless clean without sediment
Hardness (Σ Ions of alkaline earth)
≤0.02 mmol/l
Compliance with the above data should be tested in accordance with
acknowledged analytical methods, by an authorized laboratory.