Page 30
10.3. Replacing the HEPA Air Filter
Caution!
Before proceeding, make sure that the electric cord is disconnected
and there is no pressure in the autoclave.
The HEPA filter is located on the back of the autoclave. (See Rear View (7)).
1.
Pull out the filter cover. The filter cover is held in place by plastic tabs. Rotate
the cover until the tabs release, the cover will come off and the filter will come
out.
2.
The filter and cover are pressed together. Place two fingers between the filter
and the cover and while supporting the cover pull the filter and cover apart.
3.
Insert a new filter by pressing it into the filter seat.
4.
Replace the filter cover.
Note:
make sure that the arrow on the filter body points inwards, toward the chamber. Make
sure that you don’t bend the filter pipe when reattaching the cover.
Note:
It is recommended to replace the HEPA filter, every 6 months or after 1000 cycles (whichever
is the shorter period).
11.
TROUBLESHOOTING
Only technical personnel having proper qualifications and holding technical
documentation (including a technician manual) and adequate information are authorized
to service the apparatus.
Problem/
Error
Message
Message / Symbol Description
Corrective Action
Display is not
activated
The main switch is in the off position.
The power cord is not connected properly to
the machine and the power source.
There is no electrical power in the main
Turn the main switch on.
Make sure the power cord is
properly connected to the machine
and the power source.
Cap
Teflon
gasket
Strainer
element
Strainer
Housing