Page 21
Door Action/ Door
Closing
This message is displayed
while the door locking
mechanism is locking the
door.
Wait until the door locking
is completed.
Door Action/Door
Opening
This message is displayed
while the door locking
mechanism is unlocking the
door.
Wait until the door
unlocking is completed.
This message is displayed if
the electrode in the chamber
senses water.
DO NOT open the door,
water will spill out. Run a
new cycle to drain the
chamber (see 10.2)
In 2340EA-D, EKA-D,
2840EA-D, EKA-D, this
symbol is displayed when
there is no water in the
mineral-free water reservoir.
Pour water in the front
funnel until it reaches the
full level.
Supply Water
distiller error
In 3850EA-D, 3870EA-D, this
message is displayed when
there is no water in the
mineral-free water reservoir.
Fill the water reservoir until
it reaches the full level.
"Cycle Ended"
This message is displayed
when the cycle ended
successfully.
Press START/STOP in
order to perform a new
cycle.
This symbol is displayed
when Cycle by Clock mode is
active.
Enter the main menu as
described in this manual
to change the time or to
cancel this option.
"Cycle by clock"
is active
This message is displayed if
the user presses
START/STOP key while the
"start cycle by clock" mode is
active. Starting another cycle
is not allowed.
Enter the main menu as
described in this manual
to change the time or to
cancel this option.
System Ready
The system is ready to start
the cycle.
Press ^ to open
This message appears in
3850EA-D, 3870EA-D when
the door is locked but the
user may open it.
Press the Up button to
open the door
Press any button
to wake up
Screen saver is active
Press any button to stop
the screen saver.
"Atmospheric
pressure not set"
This message is displayed in
order to set the atmospheric
pressure by opening the door
for 2 minutes.
Open the door for 2
minutes in order to set the
Atmospheric pressure.