Page 23
Air-removal pulses
Heating by actuation of electrical heaters until the sterilization
temperature is reached.
Sterilization temperature is maintained constant during the sterilization
time.
Fast exhaust, steam is exhausted out of the chamber at a fast rate until
pressure decreases to ambient pressure.
For Wrapped 134 and Wrapped 121: drying by heating of chamber and
air circulation to remove leftover moisture from the instruments and
wraps. Drying time is changeable by the operator (“drying time”
parameter).
6. INSTALLATION INSTRUCTION
6.1. Placing
Caution!
The sterilizer must be placed on a rigid and leveled surface.
The stand must be able to withstand the load of the device
and loaded material. See 3.2 for unit specifications.
Make sure while placing the autoclave, to leave space around the machine, to
give the technician access to service the machine. Keep the back and the
right side of the autoclave approximately 50mm away from the wall to allow for
ventilation.
At the time of installation or anytime the unit is relocated the atmospheric
pressure parameter needs to be reset. (See 9.4).
Connect the power cord to the socket on the rear side of the autoclave; plug it
into the supply outlet. The autoclave must be connected to a properly
grounded outlet.
6.2. Setup
1.
Make sure all the feet are on the autoclave and none of them has been
lost
2.
Position the autoclave on the counter
3.
Fill the reservoir with mineral-free water (see 6.4)
4.
The unit is ready to operate.
6.3. Lifting and Carrying
Cautions!
Before moving the autoclave, Make sure that the electric
cord is disconnected from the power, and there is no
pressure in the chamber.
1. Disconnect the power supply cord.
2. Drain the water from the reservoir and the vessel: see 10.2
To avoid injuries, lifting and carrying should be done by two
people.
Do not drop the device!