Page 33
3.
The time is displayed in the form “0000” min.
4.
To increase or decrease the digits, press the UP or DOWN keys.
5.
After changing the value move the cursor to Set by pressing the
START/STOP.
6.
When Set is blinking, press the UP or DOWN keys in order to confirm
changes and return to the previous screen.
9.4. Maintenance
Reset Atmospheric Pressure
1.
This subdirectory allows resetting the atmospheric pressure.
2.
On the Main Menu, choose Maintenance\ Reset Atmospheric Pressure.
3.
The following message appears on the screen: "Reset done! New
value will be set after door is open for 2 minutes and the temperature is
less than 045.0 °C".
10.MAINTENANCE INSTRUCTIONS
10.1. Preventive and Scheduled Maintenance
The maintenance operations described in this chapter need to be followed as
indicated to keep the device in good working condition.
The instructions that follow can easily be carried out by the operating
personnel and do not require a service technician.
Note:
Technician manual describes the maintenance operations required from
qualified technician, every two months and once a year.
Should the need arise, technical assistance or a service technician can be
requested by either calling your dealer.
Daily
Clean door gasket with a mild detergent, water and a soft cloth or
sponge. The gasket should be clean and smooth.
Weekly by the operator
Clean the water sensor in the rear of the chamber with a damp cloth or
sponge. Cleaning the dirt off the sides of the sensor is more important
than the tip. (see 10.4)
Clean and descale the chamber, copper tubes, and the reservoir using
Chamber Brite (see the procedure below).
Drain the water reservoir and replace water.
Periodically by the operator
Once per month clean the water outlet strainer (see 10.3). Cleaning
frequency may be reduced according to experience.
Every 3 months check the door gasket for any signs of physical damage and
ask the technician to replace it if required.