Replacing the Locking Device
Caution:
Before starting, verify that there is no pressure in the autoclave
chamber.
1. Remove the security ring (9) using a special tool.
2. Remove pin (6).
3. Remove locking device. Take care not to lose the Teflon disk (10).
4. Reassemble the new locking device.
5. Insert the pin (6).
6. Reassemble the security ring (9).
CLOSING DEVICE
No.
Description
No.
Description
1
Bushing
6
Door locking device pin
2
Door tightening bolt
assembly
7
Bakelite handle
3
Locking screw housing
8
Closing bridge “c” clip
4
Locking base
9
Cotter pin
5
Locking housing axis
10
Teflon disk