WATER QUALITY
5.1
Water for Generating Steam
The distilled or mineral – free water supplied to the sterlizer shall be
according to the table below:
Physical Characteristics and Maximum acceptable contaminants
levels in steam for sterlizers
(According to EN 13060:2004).
Element
Condensate – allowable content
Silicium oxide. SiO
2
≤0.1 mg/kg
Iron
≤0.1 mg/kg
Cadmium
≤0.005 mg/kg
Lead
≤ 0.05 mg/kg
Rest of metals except iron,
cadmium, lead
≤0.1 mg/kg
Chloride (Cl)
≤0.1 mg/kg
Phosphate (P
2
O
5
)
≤0.1 mg/kg
Conductivity (at 20°C)
≤3 μs/cm
pH value (degree of acidity)
5 to 7
Appearance
Colourless clean without sediment
Hardness (Σ Ions of alkaline
earth)
≤0.02 mmol/l
Compliance with the above data should be tested in accordance with
acknowledged analytical methods, by an authorized laboratory.
Attention:
We recommend testing the water quality once a month. The use of
water that does not comply with the table above may have severe
impact on the working life of the sterilizer and can invalidate the
manufacturer’s guarantee.
5.2
Reverse Osmosis
A Reverse Osmosis system may be used to improve the quality of the
water used to generate steam in the autoclave chamber. The use of
mineral free will contribute to better performance and longer life of the
autoclave.
Summary of Contents for 1730 M
Page 2: ......
Page 11: ...9 REAR VIEW...
Page 34: ...32 GENERAL VIEW OF VESSEL DOOR AND ACCESSORIES Door Closing device Chamber...
Page 35: ...33 AUTOCLAVE COVER...
Page 37: ...35 MULTI PURPOSE VALVE ASSEMBLY Model 1730 Models 2340 2540 3140 3850 3870...
Page 46: ...44 DRAWING OF ELECTRICAL SYSTEM OF TABLE AUTOCLAVE MODELS 1730M MK...
Page 47: ...45 DRAWING OF ELECTRICAL SYSTEM OF TABLE AUTOCLAVE MODELS 2340 2540 M MK...
Page 48: ...46 DRAWING OF ELECTRIC SYSTEM OF TABLE AUTOCLAVE MODEL 3140 M...
Page 49: ...47 DRAWING OF ELECTRIC SYSTEM OF TABLE AUTOCLAVE MODELS 3850 M...
Page 50: ...48 DRAWING OF ELECTRICAL SYSTEM OF TABLE AUTOCLAVE MODELS 3870 M...
Page 51: ...49 PIPING DIAGRAM TABLE TOP AUTOCLAVE MODELS M AND MK...