Replacing the circuit breaker
Caution!
Before starting, disconnect the instrument from the power source.
1. Remove the autoclave cover (see para. 6.2 “Dismantling the Outer
Covers of the Autoclave
”
).
2. Disconnect the wires from the circuit breaker.
3. Remove the four screws connecting the circuit breaker to the panel (1).
4. Replace the circuit breaker with a new one.
5. Reconnect the electrical wires.
6. Reassemble the cover.
7. Turn on the autoclave and verify it operates correctly.
8. Move the circuit breaker’s lever to the “tripped” position and verify
that the autoclave turns off.
Make sure that the correct circuit breaker is installed as marked in
the table below!
1730
2340
2540
3140
3850 3870
M
Model
Voltage
M
MK
MK-V
M
MK
M
MK
standard special
M
M
1ph, 120V,
50/60 Hz
15 A 15 A 15 A — 15 A —
—
—
—
—
1ph, 230V,
50/60 Hz
10 A 10 A 10 A 15 A 10 A 15 A 10 A 15 A 15 A 15 A
Circuit breaker lever
Summary of Contents for 1730 M
Page 2: ......
Page 11: ...9 REAR VIEW...
Page 34: ...32 GENERAL VIEW OF VESSEL DOOR AND ACCESSORIES Door Closing device Chamber...
Page 35: ...33 AUTOCLAVE COVER...
Page 37: ...35 MULTI PURPOSE VALVE ASSEMBLY Model 1730 Models 2340 2540 3140 3850 3870...
Page 46: ...44 DRAWING OF ELECTRICAL SYSTEM OF TABLE AUTOCLAVE MODELS 1730M MK...
Page 47: ...45 DRAWING OF ELECTRICAL SYSTEM OF TABLE AUTOCLAVE MODELS 2340 2540 M MK...
Page 48: ...46 DRAWING OF ELECTRIC SYSTEM OF TABLE AUTOCLAVE MODEL 3140 M...
Page 49: ...47 DRAWING OF ELECTRIC SYSTEM OF TABLE AUTOCLAVE MODELS 3850 M...
Page 50: ...48 DRAWING OF ELECTRICAL SYSTEM OF TABLE AUTOCLAVE MODELS 3870 M...
Page 51: ...49 PIPING DIAGRAM TABLE TOP AUTOCLAVE MODELS M AND MK...