16
6.8
Cut-Off Thermostat
This thermostat cuts out power to the autoclave, in the event that all
other safety systems do not function. For example: if the operator
forgets to fill the chamber with water, and starts the sterilization cycle,
the chamber will heat up and activate the cut-out thermostat. In order to
restart the operation, press the RESET button. If the autoclave is
operated according to the instructions, and the thermostat cuts off, a
replace the thermostat.
The thermostat has been calibrated by the manufacturer of the
autoclave.
Do not attempt to re-calibrate it
Cut-Off Thermostat
Reset Button