13
6.4
Replacing the Safety Valve
Caution
Before starting, be sure that the electric cord is disconnected and that
there is no pressure in the autoclave.
Note:
These instructions are valid for both, CE-marked and ASME type
safety valves.
1. Remove the water reservoir cover.
2. Unscrew the safety valve and remove it from the safety valve base.
3. Replace it with a new safety valve (ensure the safety valve is an
original one!)
4. Test all autoclave.
ASME approved Type
CE marked Type