![TSI Instruments EVM SERIES User Manual Download Page 48](http://html.mh-extra.com/html/tsi-instruments/evm-series/evm-series_user-manual_1178251048.webp)
Setting up Parameters via the Keypad
Unlock run and/or setup
EVM Series User Manual
38
Unlock run and/or setup
When in secure run mode or secure setup mode, a quick press of the lock/unlock soft key and entering
in the password will unlock the instrument. (See steps 5 and 6 above to for more information on entering
in the password.)
NOTE:
If the entered password to unlock the instrument was forgotten, a default TSI password can
be used in lieu of the entered password, “9157”.
Auto-run
The EVM has four programmable auto-run settings including a timed-run, a date, a day of week (DOW),
and a disabled mode.
The EVM is shipped with the auto-run set to the disabled mode. Each mode has more than one setting
and is denoted by the “
+
” (plus) or “
–
”(minus) soft key symbols. When an alarm is activated (or
programmed), a “
+
” (plus) sign appears. The default alarm mode is set to an off mode and is denoted
via a minus symbol (i.e., “
–
”).
Auto-run and timed-run mode
With timed-run, you manually start the run (via the run key), and it will automatically stop based on the
programmed duration.
Setting auto-run with timed-run mode
1. From the
Start
screen, select and open the
Setup
screen.
2.
Select
Auto-Run
by pressing
▲
/
▼
Arrows
and press
Enter
.
3.
In the
Auto-Run
screen, press enter to select
Timed-Run
. If it is not displayed, press
▲
/
▼
Arrows
repeatedly until selected.
4.
Next, press the -
ONE
soft key or select
View/Set Parameters
.
Auto-Run and Timed Run setting
Select
Timed-run
by pressing Enter and then
press Up/Down arrows to move through the menu.
Timed Run soft key
Select first soft key to access the Timed run screen
Figure 3-22: Selecting auto-run with timed-run mode