3-2
Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938
Preparing for Shipping
To prepare the Electrostatic Classifier for moving or shipping, follow these
instructions:
1. Power off the classifier and disconnect the power cord from the power
supply.
2. If you are using a DMA, follow these instructions to remove it:
a. Remove the tubing from the inlets/outlets on the classifier.
b. Disconnect the high-voltage power supply.
c. Cap the fittings with the vinyl end caps provided with the original
packaging materials.
d. Depress the lever on the classifier base and release the DMA.
3. If you are using a Model 3088 Neutralizer, follow these instructions to
remove it:
a. Use a Phillips-head screwdriver to unscrew the two truss-head
screws holding the neutralizer in place.
b. Lift out the neutralizer column.
Summary of Contents for 3082
Page 2: ......
Page 175: ...Maintenance Service and Troubleshooting 6 33 This page intentionally left blank ...
Page 176: ......
Page 222: ......
Page 224: ......