Maintenance, Service, and Troubleshooting
6-25
Performing User Recalibrations and Checks
Check or recalibrate the sheath flow as part of your routing maintenance to
ensure ±2 percent accuracy of sheath flow rate. Full details of the sheath
flow calibration are given in
Chapter 5, Performing a User Calibration
The differential pressure sensor has some sensitivity to temperature
therefore the sensor be zeroed from time-to-time as well as before you
perform an impactor calibration.
Impactor loading can influence the accuracy of the impactor flow
measurement; therefore, check or recalibrate the impactor flow as part of
your routine maintenance. Full details of the sheath flow calibration are
given in
Chapter 5, Performing a User Calibration
Upgrading Firmware
Check the TSI web site periodically for upgrades to the Electrostatic
Classifier firmware and CPC firmware. Firmware upgrades are contained in
zip files available for download.
To find a specific firmware upgrade file, follow these instructions:
2. From the
Support
tab >
TSI Software and Firmware
>
Software and
Firmware Wizard
.
3. Follow the onscreen instructions to find the correct firmware upgrade.
Summary of Contents for 3082
Page 2: ......
Page 175: ...Maintenance Service and Troubleshooting 6 33 This page intentionally left blank ...
Page 176: ......
Page 222: ......
Page 224: ......