Instrument Operation
5-43
2. Disconnect the
Sheath Flow +
tubing from the DMA and connect it to
a reference flowmeter with a flow accuracy of
±
2% or better over the
range of 2 to 30 L/min, such as a Gilibrator
®
meter.
3. Connect the
Sheath Flow
–
port on the classifier to the
Sheath Flow
–
port on the DMA if not already connected.
4. From the classifier
Setup
screen navigate to
Calibration
>
User
Calibration
. Select the
Sheath Flow
tab.
Figure 5-28
Electrostatic Classifier User Calibration Screen
–Sheath Flow Tab
5. To check calibration at a specific flow rate, set the sheath flow to the
flow rate of interest by pressing the corresponding Setpoint button. The
Raw
and
Classifier
data fields populate, the
Reference
cell for that
row is activated, and an onscreen keypad appears.