5-8
Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938
To set the Sheath Flow on the main display screen, follow these
instructions:
1. Using your finger or a stylus, touch the power button. The power button
allows you to toggle between
On
and
Off
. If you toggle Off and then
On, the sheath flow returns to the previous set point.
Figure 5-3
Setting Sheath Flow
2. Touch the
Sheath Flow (L/min)
field.
3. Using the onscreen keypad, enter the flow rate to change the setpoint.
The current setpoint is displayed near the upper right of the Sheath
Flow field.
To set the Particle Diameter on the main display screen, follow these
instructions:
1. Using your finger or a stylus, touch the
Particle Diameter (nm)
field.
Figure 5-4
Setting Particle Diameter
2. Using the onscreen keypad, enter the desired particle size.
Note
: The particle size must be between the Min and Max (Min:5.6
Max: 238.4 in the example above) displayed beneath the size
field. The current setpoint is displayed near the upper right of
the Particle Diameter field.