SRS and XRS EID Stick Reader
Issue 2, 02/2013
27
5
Replace the case gasket taking care
that is is placed in the groove evenly and that it is located
correctly around holes and pins. The interface with the tube gasket is critical
Figure 5 Interface between the case gasket and tube gasket
6
Reconnect the FFC connecting the main module and the display module. See
circuit (FFC) cables and connectors
on
7
Close the case. See
Replacing the tube gasket
1
Open the case. See page 24.
2
Remove the main module and FRP tube assembly from the bottom case. See page 28.
3
Remove the tube gasket by sliding it off the FRP tube.
4
Slide the new tube gasket back over the FRP tube. Note that it must be orientated as shown in
Make sure that the location lugs are engaged in the holes in the tube.
5
Refit the main module and FRP tube assembly in the bottom case. See page 28.
6
Close the case. See
Replacing the battery
1
Unscrew the battery compartment cover and remove the battery.
2
Write the installation date on the new battery with a permanent marker.
3
Insert the battery with the metal battery terminals inserted first towards the main module.
4
Replace the battery compartment cover.
5
Check that the stick reader operates and place on charge for 3 hours.