Model 4590
5–21
To access the
Drift Test
function, press
〈
.
〉〈
0
〉
at the
Setup
menu.
The gauge displays the
Drift Test
menu:
To select a menu option, press the corresponding numeric key
or press
〈
ESC
〉
to return to the
Setup
menu.
Taking a Drift Test
To take a new drift test, press
〈
1
〉
at the
Drift Test
menu. Place
the gauge on a level surface with the source rod in the
SAFE
(shielded) position, and then press
〈
ENTER/START
〉
.
NOTE
If it has been less than three hours or more than eight
hours since the last stat test, the gauge displays an
error message. Press
〈
ENTER/START
〉
to return to the
Setup
menu.
The gauge displays the progress of the drift test (five 4-minute
counts). Upon completion of the counts, the gauge displays the
results of the density drift test, including an indication of
whether the test passed or failed:
Press
〈
ENTER/START
〉
to view the results of the moisture drift
test. Follow the prompts to scroll through the results.
1. Take Drift Test
2. Review Drift Test
3. Print Drift Test
<ESC> to Exit
Density Drift Test
Avg. Counts: ####
R: #.### (Pass)
<ENTER> for Moist.
Summary of Contents for EGauge 4590
Page 9: ......
Page 18: ...xvi NOTES...
Page 24: ...1 6 Figure 1 4590 Gauge and Standard Accessories...
Page 26: ...1 8...
Page 70: ......
Page 94: ...6 4...
Page 104: ......
Page 164: ......
Page 168: ......
Page 171: ...Model 4590 F 3 Standard Count Log Gauge Serial Number Date DS BGD Date DS BGD...
Page 174: ......
Page 186: ...Notes NOTES...