4–18
If conditions have changed and you want to perform a
new background count, answer
〈
NO
〉
to the question
and the following message displays:
a
a
.
.
If you have opted to take a background count, lower the
handle to the first position and press
〈
ENTER
〉
.
The
gauge takes a count and displays the following:
When the background count is finished, the display
shows:
b
b
.
.
Press
〈
ENTER
〉
to continue.
Taking a Moisture Measurement with the
Moisture Probe
In order to use the Moisture Probe, the Moisture Input must be
set to External Sensor (default setting). See page 5–8.
1
1
.
.
Turn on the probe. The green light illuminates. The blue
light on the probe remains solid when it has connected to
the Bluetooth in the gauge.
Getting background
Count Please wait
15s
With Gauge on
Test Site & Source
Rod in Background Pos.
Press <START>
BGD = xxxx
Press <ENTER>