4–8
source rod is in the background position and starts the
second step of the background count.
If the source rod is not in the background position, and if
automatic depth mode is enabled (default setting), it
displays:
9
9
.
.
When count is complete, raise handle to
SAFE
position.
1
1
0
0
.
.
The gauge checks whether the new standard count meets
the required trend and displays:
Message 1:
Message 2:
1
1
1
1
.
.
If count is acceptable, the gauge performs the following:
♦
Use the Destiny Standard -Background
(DS-BGD)
count as the
Active Standard Count.
♦
Assign filenames and archives the acquired data.
♦
Display filenames of this data.
♦
Use the
BGD
count as the Active Background.
New standard count x %
different than the
expected count.
Place the source rod to
BGD position and press
ENTER key.
New standard count
accepted
Do you want to accept
the new count?
Yes/No