NOTE
The source rod should always be in the SAFE position
when the gauge is not in use.
Instructions for Taking a Standard Count
The standard count consists of 2 separate 2-minute counts. The first count
occurs with the handle in SAFE position.
During the second part the handle is in background position and the source
rod end protrudes about 1 ½ inches into the prepared hole. This is also
different than other nuclear density gauges.
1
1
.
.
Prepare the test site as instructed on page 4–2.
2
2
.
.
Take a standard count by pressing the
〈
STD
〉
key.
3
3
.
.
Press
〈
YES
〉
to start the standard count:
4
4
.
.
Make sure the gauge is on the material to be measured and
ensure that the source rod in the
SAFE
position.
NOTE
If not in correct position this warning displays:
If you see this message, place the source rod in the
SAFE position and press <ENTER>.
4–6
Place Gauge on
Test Site & Source
Rod in SAFE Pos.
Press <START>
Standard Count
DS= xxxx BGD= xxxx
Take New Count?
<YES> or <NO>
Source Rod
NOT in Safe
Position
Press <ENTER>