GU_TRIDENT_V_1_20_EN
40/47
8.3
DETAILLED DESCRIPTION
During a measurement program, TRIDENT H acquires the input signals : 3 voltage phases and 3 current
phases. From these values, it calculates the amplitudes of the 40 first harmonics on voltages (simple or line)
and currents every second, as well as the THDs and instantaneous TDDs for each phase. These
measurements are then integrated over the period configured by the user in order to obtain the mean values.
The mean values of the first 24 harmonics (nos. 2 to 25) are compared individually to the programmed
thresholds. When there is a threshold crossing, the corresponding counter is incremented. There is a total of
144 counters (24 counters for each of the Us and Is).
In the same way, the mean values of the THDs and TDDs are compared to the programmed thresholds, and
the corresponding counters are updated. There are 3 counters for THDs and 3 counters for TDDs.
The mean values of the 3 THDs and the 3 TDDs are stored in recorder which capacity depends on the chosen
integration period :
28 days with a period of 10mn,
a little less than 3 days with a period of 1mn.
Values of the 6 signals are also stored for 3 harmonics chosen by the user. The memory capacity for the
harmonic depends on the integration period chosen :
8 days with a period of 10mn,
more than 19 hours with a period of 1mn.
All these memories are circular, i.e. when the memory is full, the oldest records are replaced by most recent.
Finally, TRIDENT H integrates the mean values of harmonics and the THDs and TDDs over 1 week,
calculates and memorizes this weekly mean. Thus the weekly mean of the first 4 weeks of the program can
be obtained.
8.4
INSTALLATION and START-UP
See chapter 4.
8.5
DETAILS OF USE AND DESCRIPTION
8.5.1
TRIDENT menu organisation
The TRIDENT menu follows the arborescence below when the device is not measuring :
Welcome
Not recording
Direct measurements
Consultation of the instantaneous values (2 pages)
Configuration
Configuration of the next
program
Confirmation
Program Parameters (4 pages)
Time & Date of start & finish
communication
Parameters
Setting the
internal clock
Consultation
Consultation of the
program parameters
(3 pages)
When a program is running the
configuration
menu is no longer accessible and it is replaced by the
ability to stop the running program:
Welcome screen (program running)
Direct Measurements
Program Stop
Consultation