![Trident TRIDENT 1U1I User Manual Download Page 20](http://html.mh-extra.com/html/trident/trident-1u1i/trident-1u1i_user-manual_1162193020.webp)
GU_TRIDENT_V_1_20_EN
20/47
Date & time of start & finish of the measurement program :
C O N F I G . P R O G R A M 7 / 7
S 0 1 / 0 1 / 2 0 0 0 0 0 : 0 0 : 0 0
E 0 1 / 0 1 / 2 0 0 0 0 0 : 0 0 : 0 0
The hour and the date of the start (S) and end (E) of the program must be entered. By default these
are the current time.
If they are not modified, the program will start immediately after the last field on this page has been
validated.
If the end date & time is identical or previous to that of the beginning, the program is of unlimited
length and will end only when terminated by the operator.
If the start date is before the present date, the program will start immediately after the last field has
been validated.
If the start date is later than the current date, TRIDENT will wait until this date to start the program.
The device can be turned off and moved, this parameter remains in the memory.