Version: TCM_0450_E80_M_EN_150209_E007
80
Listings
7.3.2.
Technical Data TCE 8000 Transmitter
General
Display
back-lit LCD screen, 132 x 32 dot
Supply Voltage
24 V DC, ±20 % or 90
… 264 V (version dependent)
Programming
via front keyboard or
Windows-based TRICOR configurator program (Modbus)
Interface
RS 485 (Modbus RTU), Option HART
®
, Foundation Fieldbus
®
other options on request
EMC
According to EN 61000-6-4 and EN 61000-6-2
Power Consumption
max. 6 W
Dimensions
see chapter 7.3.4
Electrical Connections
cage clamp terminals
Cable Glands
or 7 - 13 mm cables
Housing Material
Aluminum diecast (option: 1.4404/AISI 316L)
Protection class
IP65 (IP66/IP67 on request)
Weight:
with Aluminum diecast housing: 3.8 kg/8.3 lb with 3 m cable
with 1.4404/AISI 316L housing:
6.15 kg/13.56 lb
Temperature
ambient: -40 °F ... +158 °F) (-40 °C ... +70 °C)
storage and transport: -40 °F ... +176 °F (-40 °C ... +80 °C)
Analog Outputs
Current Outputs (2x)
4 ... 20 mA passive, two-wire, isolated
Resolution
14 bit
Linearity
±0.05 % of full scale
Temperature Drift
0.05 % per 10 K
Load
< 620
Ω (at 24 V supply)
Output Value
programmable: flow, total, density, temperature
Pulse Output
Frequency Range
0.5
-
10,000 Hz
Output Signal
active push pull output for flow rate
Status In- and Output
Status Output
push pull programmable
(in TOTAL mode: 0.5 - 100 Hz; in FREQUENCY mode: 0.5 - 10,000 Hz)
Control Input
programmable
Analog Input (option)
Input Type
4-20 mA active for two-wire passive pressure sensor
Resolution
12 bit
Linearity
±0.05 % of full scale
Temperature Drift
0.05 % per 10 K
Supply Voltage
> 20 V (at 20 mA sensor current)