Manual Operation
FREQUENCY:
A frequency proportional to the current flow rate is generated. Frequencies between 0.5 Hz and 10 kHz can be
generated in this mode.
PHASE SHIFT:
Prerequisite: the FREQ-OUT has been programmed for TOTAL COUNT. As soon as
“TOTAL” has increased by
the chosen amount, another pulse is generated on FREQ OUT. To have a frequency ratio of 50% on the output,
the state of the output changes respectively after half the amount. A pulse offset by 90° is also generated on
CTRL OUT after each pulse on FREQ-OUT.
The maximum output frequency in this operating mode is about 50 Hz.
FAULT
CTRL OUT MENU
FAULT
UP DOWN EXIT
P
SELECT ACTIVE STATE
ACTIVE HIGH
UP DOWN EXIT
Use the keys
“UP” and “DOWN” to select the desired value and confirm with “P” or skip with “EXIT”.
The display returns to the SETUP IN/OUTPUTS menu.
FLOW DIR
CTRL OUT MENU
FLOW DIR
UP DOWN EXIT
P
SELECT ACTIVE STATE
ACTIVE HIGH
UP DOWN EXIT
Use the keys
“UP” and “DOWN” to select the desired value and confirm with “P” or skip with “EXIT”.
The display returns to the SETUP IN/OUTPUTS menu.
BATCH
CTRL OUT MENU
BATCH
UP DOWN EXIT
P
BATCH VALUE
0.50000 KG
RIGHT UP EXIT
Use the keys
“RIGHT” and “UP” to select the desired step value and confirm with “P” or skip with “EXIT”.
The display shows
SELECT ACTIVE STATE
ACTIVE HIGH
UP DOWN EXIT
Use the keys
“UP” and “DOWN” to select the desired value and confirm with “P” or skip with “EXIT”.
The display returns to the SETUP IN/OUTPUTS menu.