Version: TCM_0450_E80_M_EN_150209_E007
55
Manual Operation
4.5.7.
SETUP FILTER Menu
In the submenu SETUP/FILTER the filters of TCE can be configured.
SETUP MENU
FILTER
UP DOWN EXIT
P
FILTER MENU
FLOW
UP DOWN EXIT
Use the keys
“UP” and “DOWN” to select the desired submenu and confirm with “P” or skip with “EXIT”.
4.5.7.1.
FLOW-FILTER Menu
In the submenu FLOW-FILTER the time constant for the flow filter can be set.
The time constant is the time the output needs after a jump from x to 0 to go to x/e = x/2.72.
A rough relation between the time and the filtered flow value after a jump is
Elapsed time
Remaining error
(% of the step)
1 *
t
30
2 *
t
10
3 *
t
3
4 *
t
1
FILTER MENU
FLOW
UP DOWN EXIT
P
FLOW FILTER 0-99.9 SEC
1.00000
RIGHT UP EXIT
Use the keys
“RIGHT” and “UP” to select the desired value and confirm with “P” or skip with “EXIT”.
The display returns to the SETUP FILTER menu.