19
Getting Started
2.4.
Quick Start
WARNING!
As for safety and accuracy reasons many precautions must be taken, read chapter 3 carefully before installing
the TCM!
In case the TCM has only to be operated without flow for testing or learning purpose, at least the following
connections have to be made (see chapter 3.2):
Connect the TCE to the TCM
Connect the supply voltage
The various inputs and outputs as well as the interface may be connected as well, if those features
are required.
WARNING!
If the TRICOR Coriolis Mass Flow Meter is connected to a bigger system, for your personal safety connect the
protective ground as well!
WARNING!
In hazardous areas it is not allowed to operate the TRICOR Coriolis Mass Flow Meter without proper wiring in
accordance with chapter 3.2.5 and with the housing not properly closed!
2.4.1.
First Operation
Make sure that all mechanical and electrical connections are made properly.
Switch on the power supply. The LED
“OK” will flash green.
After the power up sequence the display shows the preselected values (ex factory:
“Flow” and “Batch”)
Switch on the flow. The value indicated in the display should be positive.
In case of an error the LED
“ERR” will flash red.
As soon as the TCM has reached the operating temperature, make the zero point calibration (see chapter 4.2.1
for detailed information):
Switch off the flow
Wait until the flow is zero
Start the zero offset calibration in the ZERO OFFSET menu
Wait until the offset procedure is finished
Switch on the flow again
The display can be altered by pressing the pushbutton
“Display”.
The internal device status can be viewed by pressing the pushbutton
“Info”.
If the function is activated, the BATCH reading can be reset to zero by pressing the pushbutton
“Reset”.
To open the control menu press button
“P” for three seconds.