JJ Calibrations, Inc.
7007 SE Lake Rd
Portland, OR 97267-2105
Phone 503.786.3005
FAX 503.786.2994
Certificate of Calibration
Certificate Number:
692254
PO:
Property #:
Serial #:
User:
Model:
Description:
Make:
Accuracy:
*Recommended Due:
As Received:
As Returned:
Action Taken:
Order Date:
Procedure:
Calibrated on:
Technician:
°C
% RH
Environment:
Department:
B729400181
N/A
Mettler Toledo
07/11/2019
Within Tolerance
01/11/2019
19
123
Calibrated
±0.0005g
DCN 500887
N/A
181203
01/11/2019
OMNI-00637
Analytical Scale, 120g
Within Tolerance
43
O n S i t e
MS104TS/00
Authorized By:
N/A
0723.01
256A
Rice Lake
W0133K
Mass Set,
05/30/2019 660578
Std ID Manufacturer
Model
Nomenclature
Due Date
Trace ID
Standards Used
Parameter
Unit
UUT Uncertainty
Measurement Description
Range
Measurement Data
Before/After
Accredited =
ü
Min
Max
Reference
*Error
Force
g
5.7E-04
10.00000
10.0000
0.0000
ü
9.9995 10.0005
g
g
5.7E-04
30.00000
30.0000
0.0000
ü
29.9995 30.0005
g
g
5.7E-04
60.00000
59.9998
0.0002
ü
59.9995 60.0005
g
g
5.7E-04
90.00000
89.9999
0.0001
ü
89.9995 90.0005
g
g
5.7E-04
120.00000
119.9998
0.0002
ü
119.9995 120.0005
g
Remarks:
* Many factors may cause the unit to drift out of calibration before the recommended due date. Any reported error is the absolute value between the reference and the unit.
Omni-Test Laboratories
13327 NE Airport Way
Portland, OR 97230
Calibration
Uncertainties include the effects of the unit.
*
*
Inspector
Rev #
Issued
15
01/14/2019
Reviewer
3
JJ Calibrations, Inc. certifies that this instrument has been calibrated in accordance with the JJ Calibrations Quality Assurance Manual with the stated procedure using
standards that are traceable to the National Institute of Standards and Technology (NIST), or other National Measurement Institutes (NMI's), or by using natural
physical constants, intrinsic standards or ratio calibration techniques. The quality system and this certificate are in compliance with ANSI/NCSL Z540-1-1994,
ISO/IEC 17025-2005, ISO 10012-1, the ISO 9000 family and QS 9000. The expanded uncertainties of measurements for this calibration are based upon 95%
(2 sigma) confidence limits. Unless otherwise stated, a test accuracy ratio (TAR) of 4:1, if achievable, is maintained. The results reported herein apply only to the
calibration of the item described above. This report may not be reproduced, except in full, without prior written consent of JJ Calibrations, Inc.
JJ Calibrations, Inc. quality system has been assessed and accredited to ISO/IEC 17025:2005.
Page 1 of 1
Certificate:
692254
Page 135 of 186
Summary of Contents for Endeavor
Page 22: ...Page 22 of 186 ...
Page 23: ...Page 23 of 186 ...
Page 24: ...Page 24 of 186 ...
Page 34: ...Page 34 of 186 ...
Page 35: ...Page 35 of 186 ...
Page 56: ...Page 56 of 186 ...
Page 57: ...Page 57 of 186 ...
Page 67: ...Page 67 of 186 ...
Page 68: ...Page 68 of 186 ...
Page 91: ...Page 91 of 186 ...
Page 103: ...Page 103 of 186 ...
Page 104: ...Page 104 of 186 ...
Page 108: ...Page 108 of 186 ...
Page 109: ...Page 109 of 186 ...
Page 110: ...Page 110 of 186 ...
Page 111: ...Page 111 of 186 ...
Page 112: ...Page 112 of 186 ...
Page 113: ...Page 113 of 186 ...
Page 114: ...Page 114 of 186 ...
Page 116: ...Page 116 of 186 ...
Page 117: ...Page 117 of 186 ...
Page 118: ...Page 118 of 186 ...
Page 120: ...Page 120 of 186 ...
Page 121: ...Page 121 of 186 ...
Page 122: ...Page 122 of 186 ...
Page 123: ...Page 123 of 186 ...
Page 124: ...Page 124 of 186 ...
Page 127: ...7 20 18 Page 127 of 186 ...
Page 130: ...7 20 18 Page 130 of 186 ...
Page 131: ...Page 131 of 186 ...
Page 133: ...Page 133 of 186 ...
Page 136: ...Page 136 of 186 ...
Page 184: ...34 Optional Equipment Travis Industries 2 25 2019 1507 Endeavor II Listing Label ...