background image

 

 

28

29

COMPLIED STANDARDS LIST

FCC STATEMENT

Complied Standards List

EN ISO 14971:2012 / ISO 14971:2007 Medical devices - 

Application of risk management to medical devices

EN 1041:2008 Information supplied by the manufacturer of medical 

devices

EN 60601-1:2006+A1:2013/ IEC 60601-1:2005+A1:2012 Medical 

electrical equipment - Part 1: General requirements for basic safety 

and essential performance

EN 60601-1-11:2015/ IEC 60601-1-11:2015 Medical electrical 

equipment - Part 1-11: General requirements for basic safety and 

essential performance - Collateral standard: Requirements for medical 

electrical equipment and medical electrical systems used in the home 

healthcare environment

EN 60601-1-2:2015/ IEC 60601-1-2:2014 Medical electrical 

equipment - Part 1-2: General requirements for basic safety and 

essential performance - Collateral standard: Electromagnetic 

disturbances - Requirements and tests

EN ISO 81060-1:2012 Non-invasive sphygmomanometers - Part 1: 

Requirements and test methods for non-automated measurement type

EN 1060-3:1997+A2:2009 Non-invasive sphygmomanometers - 

Part 3: Supplementary requirements for electro-mechanical blood 

pressure measuring systems

EN 1060-4:2004 Non-invasive sphygmomanometers - Part 4: Test 

procedures to determine the overall system accuracy of automated 

non-invasive sphygmomanometers

EN 60601-1-6:2010+A1:2015/IEC 60601-1-6:2010+A1:2013 

Medical electrical equipment - Part 1-6: General requirements for basic 

safety and essential performance - Collateral standard: Usability

IEC 62366-1:2015 Medical devices - Part 1: Application of 

usability engineering to medical devices

EN 62304:2006/AC: 2008 / IEC 62304: 2006+A1:2015   Medical 

device software - Software life-cycle processes

Risk management

Labeling
User manual

General Requirements 

for Safety

Electromagnetic

compatibility

Performance

requirements

Clinical investigation

Usability

Software life-cycle 

processes

Bio-compatibility

ISO 10993-1:2009 Biological evaluation of medical devices- Part 

1: Evaluation and testing within a risk management process

ISO 10993-5:2009 Biological evaluation of medical devices - 

Part 5: Tests for in vitro cytotoxicity

ISO 10993-10:2010 Biological evaluation of medical devices - 

Part 10: Tests for irritation and skin sensitization

EN ISO 15223-1:2016 / ISO 15223-1:2016  Medical devices. 

Symbols to be used with medical device labels, labelling and 

information to be supplied. Part 1 : General requirements

IEC 80601-2-30:2009+A1:2013 Medical electrical equipment- Part 

2-30: Particular requirements for the basic safety and essential 

performance of automated non-invasive sphygmomanometers

ISO 81060-2:2013  Non-invasive sphygmomanometers - Part 2: 

Clinical validation of automated measurement type

FCC Statement

This device complies with Part 15 of the FCC Rules. Operation is subject to the following

two conditions: (1) this device may not cause harmful interference, and (2) this device must 

accept any interference received, including interference that may cause undesired operation.

Caution: The user is cautioned that changes or modifications not expressly approved by the

party responsible for compliance could void the user's authority to operate the equipment.

NOTE: This equipment has been tested and found to comply with the limits for a Class B 

digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide 

reasonable protection against harmful interference in a residential installation. This

equipment generates, uses and can radiate radio frequency energy and, if not installed and 

used in accordance with the instructions,may cause harmful interference to radio 

communications. However, there is no guarantee that interference will not occur in a 

particular installation.

If this equipment does cause harmful interference to radio or television reception, which 

can be determined by turning the equipment off and on, the user is encouraged to try to 

correct the interference by one or more of the following measures:

-- Reorient or relocate the receiving antenna.

-- Increase the separation between the equipment and receiver.

-- Connect the equipment into an outlet on a circuit different from that to which the receiver 

is connected.

-- Consult the dealer or an experienced radio/TV technician for help.

FCC Radiation Exposure Statement:

This equipment complies with FCC radiation exposure limits set forth for an uncontrolled 

environment. This transmitter must not be co-located or operating in conjunction with any 

other antenna or transmitter.

FCC ID

OU9TMB1872BS

Summary of Contents for TMB-1872-B

Page 1: ...e Monitor TMB 1872 B Please read the user manual carefully and thoroughtly so as to ensure the safe usage of this product Keep the manual well for further reference in case you have problems Guangdong...

Page 2: ...he Records INFORMATION FOR USER 21 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 23 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Irregu...

Page 3: ...rinciple Contraindications 1 The device should not be used by any person who may be suspected of or is pregnant 2 The device is not suitable for use on patients with implanted electrical devices such...

Page 4: ...oes not result in prolonged impairment of patient blood circulation When measurement please avoid compression or restriction of the connection tubing The device cannot be used with HF surgical equipme...

Page 5: ...neously Cleaning Dust environment may affect the performance of the unit Please use the soft cloth to clean the whole unit before and after use Don t use any abrasive or volatile cleaners The device d...

Page 6: ...ies 2 Installing and Replacing the Batteries CAUTION CAUTION Replace the batteries whenever the below happens The shows The display is dim The display does not light up Do not use new and used batteri...

Page 7: ...hen press SET button for 3 seconds to enter the mode for year setting 3 When you get the right year press SET button to set down and turn to next step 4 Repeat steps 2 and 3 to set the MONTH and DAY 5...

Page 8: ...connect with your smartphone on the app to continute the pair up Bluetooth Module No LS8261 RF Frequency Range 2400 MHz to 2483 5 MHz Output Power Range 8dBm Supply Voltage 1 9 3 6 V Transmitting Dist...

Page 9: ...asurements at approximately the same time on the same arm or as directed by a physician 5 6 Take the measurement in a silent room The cuff should maintain at the same level as the right atrium of the...

Page 10: ...mbol will not blink then the device will turn off 1 When the monitor is off press Start Stop button to turn on the monitor and it will finish the whole measurement Adjust the zero LCD display Inflatin...

Page 11: ...ed from the list The corresponding time is 10 38 The corresponding date is 2019 year July 1st CAUTION The date and time of the record will be shown alternately The current No is No 1 three records in...

Page 12: ...smoking When talking or moving your fingers When you want to discharge urine 1 Hold pressing MEM button for 3 seconds when the monitor is in the memory recall mode the flash display dEL ALL User ID w...

Page 13: ...usable cuff with water and never immerse the cuff in water An irregular heartbeat is detected when a heartbeat rhythm varies while the device is measuring systolic pressure and diastolic pressure Duri...

Page 14: ...his section includes a list of error messages and frequently asked questions for problems you may encounter with your blood pressure monitor If the products not operating as you think it should check...

Page 15: ...protection Protection against ingress of water Software Version A01 Pressure 5 C 40 C within 3mmHg 0 4kPa Pulse value 5 Device Classification Battery Powered Mode Internally Powered ME Equipment AC Ad...

Page 16: ...Biological evaluation of medical devices Part 10 Tests for irritation and skin sensitization EN ISO 15223 1 2016 ISO 15223 1 2016 Medical devices Symbols to be used with medical device labels labellin...

Page 17: ...hat supplies buildings used for domestic purposes Guidance and manufacturer s declaration electromagnetic immunity Immunity test 8 kV contact 15 kV air 8 kV contact 15 kV air 2 kV power supply lines l...

Page 18: ...te for the frequency of the transmitter Recommended separation distances d 0 35 d 1 2 10V m 80 Am at 1kHz 80 MHz to 800 MHz d 1 2 800 MHz to 2 7 GHz d 2 3 where P is the maximum output power rating of...

Page 19: ...960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18Hz 2 0 3 28 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217Hz 2 0 3...

Reviews: