INTRODUCTION

INTRODUCTION

LCD Display Signal

User 1

Start measurement for User 1

User 2

Start measurement for User 2

SYMBOL

DESCRIPTION

Systolic blood pressure

High blood pressure

Diastolic blood pressure Low blood pressure

EXPLANATION

Memory

mmHg

Measurement Unit of the blood pressure

Low battery

Batteries are low and need to be replaced

mmHg

Current Time

Year/Month/Day, Hour/Minute

Pulse in beats per minute

Pulse display

Blood pressure monitor is detecting an 

irregular heartbeat during measurement.

Irregular heartbeat

Blood pressure 

level indicator

Indicate the blood pressure level

(1mmHg=0.133kPa)

Indicate it is in the memory mode and 

which group of memory it is.

 Heartbeat

Blood pressure monitor is detecting a 

heartbeat during measurement.

Motion indicator

Motion may result in an inaccurate

measurement

Average value 

The average value of blood pressure

Bluetooth icon

The bluetooth icon blinks when the bluetooth is 

working

CAUTION

* The service life of the cuff may vary by the frequency of washing, skin condition, and 

storage state. The typical service life is 10000 times.

* It is recommended that the performance should be checked every 2 years and after 

maintenance and repair, by retesting at least the requirements in limits of the error of the 

cuff pressure indication and air leakage (testing at least at 50mmHg and 200mmHg).

* Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according 

to the local guidelines.

* Manufacturer will make available on request circuit diagrams, component part lists, 

descriptions, calibration instructions,etc., to assist to service personnel in parts repair.

* The plug/adapter plug pins insulate the device from the main supply. Do not position the 

device in a position where it is difficult to disconnect from the supply mains to safely 

terminate operation of ME equipment.

* The operator shall not touch output of batteries/adapter and the patient simultaneously.

* Cleaning :Dust environment may affect the performance of the unit. Please use the soft 

cloth to clean the whole unit before and after use. Don’t use any abrasive or volatile 

cleaners.

* The device doesn’t need to be calibrated within two years of reliable service.

* If you have any problems with this device, such as setting up, maintaining or using, please 

contact the SERVICE PERSONNEL of Transtek. Don’t open or repair the device by yourself 

in the event of malfunctions. The device must only be serviced, repaired and opened by 

individuals at authorized sales/service centers.

* Please report to Transtek if any unexpected operation or events occur.

* Keep the unit out of reach of infants, young children or pets to avoid inhalation or 

swallowing of small parts. It is dangerous or even fatal.

* Be careful to strangulation due to cables and hoses, particularly due to excessive length.

* At least 30 min required for ME equipment to warm from the minimum storage 

temperature between uses until it is ready for intended use. At least 30 min required for ME 

equipment to cool from the maximum storage temperature between uses until it is ready for 

intended use.

* This equipment needs to be installed and put into service in accordance with the 

information provided in the ACCOMPANYING DOCUMENTS;

* Wireless communications equipment such as wireless home network devices, mobile 

phones, cordless telephones and their base stations, walkie-talkies can affect this 

equipment and should be kept at least a distance d away from the equipment. The distance 

d is calculated by the MANUFACTURER from the 80MHz to 5.8 GHz column of Table 4 and 

Table 9 of IEC 60601-1-2:2014, as appropriate.

* Please use ACCESSORIES and detachable parts specified/ authorised by 

MANUFACTURE. Otherwise, it may cause damage to the unit or danger to the 

user/patients.

* There is no luer lock connectors are used in the construction of tubing, there is a 

possibility that they might be inadvertently connected to intravascular fluid systems, 

allowing air to be pumped into a blood vessel.

* Please use the device under the environment which was provided in the user manual. 

Summary of Contents for TMB-1872-B

Page 1: ...e Monitor TMB 1872 B Please read the user manual carefully and thoroughtly so as to ensure the safe usage of this product Keep the manual well for further reference in case you have problems Guangdong...

Page 2: ...he Records INFORMATION FOR USER 21 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 23 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Irregu...

Page 3: ...rinciple Contraindications 1 The device should not be used by any person who may be suspected of or is pregnant 2 The device is not suitable for use on patients with implanted electrical devices such...

Page 4: ...oes not result in prolonged impairment of patient blood circulation When measurement please avoid compression or restriction of the connection tubing The device cannot be used with HF surgical equipme...

Page 5: ...neously Cleaning Dust environment may affect the performance of the unit Please use the soft cloth to clean the whole unit before and after use Don t use any abrasive or volatile cleaners The device d...

Page 6: ...ies 2 Installing and Replacing the Batteries CAUTION CAUTION Replace the batteries whenever the below happens The shows The display is dim The display does not light up Do not use new and used batteri...

Page 7: ...hen press SET button for 3 seconds to enter the mode for year setting 3 When you get the right year press SET button to set down and turn to next step 4 Repeat steps 2 and 3 to set the MONTH and DAY 5...

Page 8: ...connect with your smartphone on the app to continute the pair up Bluetooth Module No LS8261 RF Frequency Range 2400 MHz to 2483 5 MHz Output Power Range 8dBm Supply Voltage 1 9 3 6 V Transmitting Dist...

Page 9: ...asurements at approximately the same time on the same arm or as directed by a physician 5 6 Take the measurement in a silent room The cuff should maintain at the same level as the right atrium of the...

Page 10: ...mbol will not blink then the device will turn off 1 When the monitor is off press Start Stop button to turn on the monitor and it will finish the whole measurement Adjust the zero LCD display Inflatin...

Page 11: ...ed from the list The corresponding time is 10 38 The corresponding date is 2019 year July 1st CAUTION The date and time of the record will be shown alternately The current No is No 1 three records in...

Page 12: ...smoking When talking or moving your fingers When you want to discharge urine 1 Hold pressing MEM button for 3 seconds when the monitor is in the memory recall mode the flash display dEL ALL User ID w...

Page 13: ...usable cuff with water and never immerse the cuff in water An irregular heartbeat is detected when a heartbeat rhythm varies while the device is measuring systolic pressure and diastolic pressure Duri...

Page 14: ...his section includes a list of error messages and frequently asked questions for problems you may encounter with your blood pressure monitor If the products not operating as you think it should check...

Page 15: ...protection Protection against ingress of water Software Version A01 Pressure 5 C 40 C within 3mmHg 0 4kPa Pulse value 5 Device Classification Battery Powered Mode Internally Powered ME Equipment AC Ad...

Page 16: ...Biological evaluation of medical devices Part 10 Tests for irritation and skin sensitization EN ISO 15223 1 2016 ISO 15223 1 2016 Medical devices Symbols to be used with medical device labels labellin...

Page 17: ...hat supplies buildings used for domestic purposes Guidance and manufacturer s declaration electromagnetic immunity Immunity test 8 kV contact 15 kV air 8 kV contact 15 kV air 2 kV power supply lines l...

Page 18: ...te for the frequency of the transmitter Recommended separation distances d 0 35 d 1 2 10V m 80 Am at 1kHz 80 MHz to 800 MHz d 1 2 800 MHz to 2 7 GHz d 2 3 where P is the maximum output power rating of...

Page 19: ...960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18Hz 2 0 3 28 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217Hz 2 0 3...