background image

32

33

EMC GUIDANCE

Table 3

Guidance and manufacturer’s declaration – electromagnetic immunity

Immunity test

Compliance 

level

IEC 60601 

Test level

Conducted RF

IEC 61000-4-6 

Radiated RF 

IEC 61000-4-3

10V/m, 80% Am

at 1kHz

150 kHz to

80 MHz: 

3 Vrms

6Vrms (in ISM 

and amateur 

radio bands)

80% Am at 1kHz 

Electromagnetic environment - guidance

NOTE 1 

At 80 MHz and 800 MHz, the higher frequency range applies. 

NOTE 2 

These guidelines may not apply in all situations. Electromagnetic propagation is affected by 

absorption and reflection from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) 

telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast 

cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to 

fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field 

strength in the location in which the device is used exceeds the applicable RF compliance level 

above, the device should be observed to verify normal operation. If abnormal performance is 

observed, additional measures may be necessary, such as re-orienting or relocating the device.

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

b

The device is intended for use in the electromagnetic environment specified below. The customer or the 

user of the device should assure that it is used in such an environment.

150 kHz to

80 MHz: 

3 Vrms

6Vrms (in ISM 

and amateur 

radio bands)

80% Am at 1kHz 

Portable and mobile RF communications 

equipment should be used no closer to any part 

of the device, including cables, than the 

recommended separation distance calculated 

from the equation appropriate for the frequency 

of the transmitter.

Recommended separation distances:

d=0.35        ;

d=1.2

10V/m, 80% Am

at 1kHz

80 MHz to 800 MHz:

d=1.2

800 MHz to 2.7 GHz:

d=2.3

where,

 P

 is the maximum 

output power rating of the 

transmitter in watts (W) 

according to the 

transmitter manufacturer, 

d

 is the recommended 

separation distance in 

meters (m).  Field 

strengths from fixed RF 

transmitters, as 

determined by an 

electromagnetic site 

survey, should be less 

than the compliance level 

in each frequency range.

Interference may occur in 

the vicinity of equipment 

marked with the following

symbol:

P

P

P

P

a

b

 

150 kHz to 80 MHz 

80 MHz to 800 MHz 

800 MHz to 2.7 GHz 

Table 4

Recommended separation distances between portable and mobile RF communications 

equipment and the device.

Rated maximum output

power of transmitter

(W) 

Separation distance according to frequency of transmitter

 (m)

0.01

0.1

1

10

100

0.12

0.12
0.38
1.2
3.8
12

0.23
0.73
2.3
7.3
23

=

d

=

d

1.2

2.3

=

d

0.38
1.2
3.8
12

3.5

The device is intended for use in an electromagnetic environment in which radiated RF disturbances are 

controlled. The customer or the user of the device can help prevent electromagnetic interference by 

maintaining a minimum distance between portable and mobile RF communications equipment 

(transmittters) and the device as recommended below, according to the maximum output power of the 

communications equipment.

For transmitters rated at a maximum output power not listed above, the recommended separation 

distance 

d

 in metres (m) can be estimated using the equation applicable to the frequency of the 

transmitter, where 

P

 is the maximum output power rating of the transmitter in watts (W) according to the 

transmitter manufacturer.

NOTE 1    At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2    These guidelines may not apply in all situations. Electromagnetic propagation is affected by 

absorption and reflection from structures, objects and people.

P

P

P

EMC GUIDANCE

Summary of Contents for TMB-1872-B

Page 1: ...e Monitor TMB 1872 B Please read the user manual carefully and thoroughtly so as to ensure the safe usage of this product Keep the manual well for further reference in case you have problems Guangdong...

Page 2: ...he Records INFORMATION FOR USER 21 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 23 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Irregu...

Page 3: ...rinciple Contraindications 1 The device should not be used by any person who may be suspected of or is pregnant 2 The device is not suitable for use on patients with implanted electrical devices such...

Page 4: ...oes not result in prolonged impairment of patient blood circulation When measurement please avoid compression or restriction of the connection tubing The device cannot be used with HF surgical equipme...

Page 5: ...neously Cleaning Dust environment may affect the performance of the unit Please use the soft cloth to clean the whole unit before and after use Don t use any abrasive or volatile cleaners The device d...

Page 6: ...ies 2 Installing and Replacing the Batteries CAUTION CAUTION Replace the batteries whenever the below happens The shows The display is dim The display does not light up Do not use new and used batteri...

Page 7: ...hen press SET button for 3 seconds to enter the mode for year setting 3 When you get the right year press SET button to set down and turn to next step 4 Repeat steps 2 and 3 to set the MONTH and DAY 5...

Page 8: ...connect with your smartphone on the app to continute the pair up Bluetooth Module No LS8261 RF Frequency Range 2400 MHz to 2483 5 MHz Output Power Range 8dBm Supply Voltage 1 9 3 6 V Transmitting Dist...

Page 9: ...asurements at approximately the same time on the same arm or as directed by a physician 5 6 Take the measurement in a silent room The cuff should maintain at the same level as the right atrium of the...

Page 10: ...mbol will not blink then the device will turn off 1 When the monitor is off press Start Stop button to turn on the monitor and it will finish the whole measurement Adjust the zero LCD display Inflatin...

Page 11: ...ed from the list The corresponding time is 10 38 The corresponding date is 2019 year July 1st CAUTION The date and time of the record will be shown alternately The current No is No 1 three records in...

Page 12: ...smoking When talking or moving your fingers When you want to discharge urine 1 Hold pressing MEM button for 3 seconds when the monitor is in the memory recall mode the flash display dEL ALL User ID w...

Page 13: ...usable cuff with water and never immerse the cuff in water An irregular heartbeat is detected when a heartbeat rhythm varies while the device is measuring systolic pressure and diastolic pressure Duri...

Page 14: ...his section includes a list of error messages and frequently asked questions for problems you may encounter with your blood pressure monitor If the products not operating as you think it should check...

Page 15: ...protection Protection against ingress of water Software Version A01 Pressure 5 C 40 C within 3mmHg 0 4kPa Pulse value 5 Device Classification Battery Powered Mode Internally Powered ME Equipment AC Ad...

Page 16: ...Biological evaluation of medical devices Part 10 Tests for irritation and skin sensitization EN ISO 15223 1 2016 ISO 15223 1 2016 Medical devices Symbols to be used with medical device labels labellin...

Page 17: ...hat supplies buildings used for domestic purposes Guidance and manufacturer s declaration electromagnetic immunity Immunity test 8 kV contact 15 kV air 8 kV contact 15 kV air 2 kV power supply lines l...

Page 18: ...te for the frequency of the transmitter Recommended separation distances d 0 35 d 1 2 10V m 80 Am at 1kHz 80 MHz to 800 MHz d 1 2 800 MHz to 2 7 GHz d 2 3 where P is the maximum output power rating of...

Page 19: ...960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18Hz 2 0 3 28 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217Hz 2 0 3...

Reviews: