6
SETUP MENU DESCRIPTIONS
This section provides more detailed descriptions of the selections found in the Setup Menu Chart.
Factory-set defaults are shown in bold; (NA) for North America and (
€
) for Europe.
NOTE: Some selections are subject to local legal metrology regulations
CODE/NAME
DESCRIPTION
SELECTION LIST
F1
Graduations
Specifies number of full-scale graduations, i.e. capacity / division.
Value should be consistent with legal regulations and environmental
limits on the useful system resolution. Pressing the ZERO key to
scroll down one level begins the sequence
Key-in
100 - 50000
5000 (NA)
3000 (€)
F2
Sampling Rate
Sets the sampling rate in Hertz (measurements per second). Use
10 Hz for most applications or 80 Hz for extra fast response time
10 (NA/€) 80
F3
Zero Track
Band
Selects the range within which the scale will automatically zero.
Note that the scale must be in standstill to automatically zero.
Selections are in display divisions (d).
0d
0.5d (NA/€)
1d
3d
5d
F4
Zero Range
Selects the range (expressed as a percentage of full scale capacity)
within which the scale may be zeroed. Note that the indicator must
be in standstill to zero the scale.
100% (NA)
1.9%
2% (€)
20%
F5
Motion Band
Sets the level at which motion is detected. If motion is not detected,
the scale can process a Print or Zero command. Maximum value
varies depending on local regulations. Expressed as scale divisions
per second (d/s). Pressing the ZERO key to scroll down one level
begins the sequence
Key-in
0.0d/s – 32.0d/s
1.0d/s (NA/€)
F6
Digital Filter
Averages weight readings to produce higher stability. Choose the
speed that works best for your application.
“FAST” = Fast “nnEd” = Medium “SLo” = Slow
FAST
nnEd (NA/€)
SLo
F7
Overload Limit
Selects the desired formula which determines the point at which the
indicator shows overload. All selections are based on the primary
unit selected in F8.
"FS" = Full scale capacity.
FS
FS + 2% (NA)
FS + 5%
FS + 1d
FS + 9d (€)
F8
Calib. Unit
Selects the primary base unit to be used in the calibration process.
Also the default unit for normal operation.
"1" = primary unit is lb. "2" = primary unit is in kg.
1 (NA)
2 (€)
F9
Display
Divisions
Determines the desired weight increments. Value should be
consistent with legal requirements.
1 (NA)
2
5 (€)
F10
Decimal Pt.
Determines location of the decimal point.
0 (NA) 0.0 (€)
0.00 0.000
0.0000 00