3
CONNECTING THE WEIGH PLATFORM
ATTENTION INSTALLERS:
It is highly recommended that you install the enclosed green/yellow cable between
the load cell cable shield and the AC ground lug on the main PCBA. Do not connect
the load cell cable shield to the green load cell input terminal.
The TI-500E-SS indicator contains a connection terminal on the main board assembly for connection to
the load cell cable. . NOTE: 6-wire connection is optional.
Load Cell Terminal (J4)
Pin No.
Wire Name
Pin No.
Wire Name
1
Do Not Use
5
- Excitation
2
- Signal
6
+ Sense
3
+ Signal
7
+ Excitation
4
- Sense