8
CODE/NAME
DESCRIPTION
SELECTION LIST
F34
Auto Print Min.
Weight
Selects the minimum weight at which the auto print function will
work if enabled. Selections are in display divisions (d). Scrolling
down with the ZERO key one level begins the procedure.
“0” = Disabled
Key-in
0 - 100
1 (NA/€)
SETUP MENU PROCEDURES
This section provides instructions for all of the Setup Menu procedures except F16 through F20 whose
procedures can be found in the Calibration section.
Fine-tune 4-20 mA output (F23)
1. While in the Setup Menu mode, scroll to "F 23", then scroll down once using the ZERO key
to enter Fine-tune menu. The indicator outputs 4 mA and displays a number.
2. While monitoring the voltage across R
L
(see Appendix E), use the right (PRINT) or left
(TARE) keys to change the displayed value until the measured voltage is exactly 1 VDC.
3. Press the SET (Net/Gross) key to save. The indicator outputs 20 mA and displays another
number.
4. While monitoring the voltage across R
L
, use the right (PRINT) or left (TARE) keys to change
the displayed value until the measured voltage is exactly 5 VDC.
5. Press the SET (Net/Gross) key to save and revert back to F23.