CONNECTING THE SERIAL I/O DEVICE 

 
The TI-500E SS model comes standard with one full duplex RS-232 serial port, designed for connection 
to a computer or a serial printer. The same port may be also used as a simplex, RS-232 port designed 
for connection to a remote display.  
 
Connection assignments for all serial RS-232 communication terminals on the TI-500E-SS are shown 
below.  

RS-232 Terminal (J3) 

 

Pin No. 

Wire Name 

 

Pin No. 

Wire Name 

TXD 

 

CTS 

RXD 

 

Ground 

RTS 

 

 

 

 

 

CONNECTING THE POWER SUPPLY 

 
The TI-500E-SS indicator ships with an AC line cord attached to the indicator. The internal power supply 
is rated for 100-240 VAC and 50/60 Hz. Simply plug the unit into a standard wall outlet.  

 

 
 

CONFIGURATION 

 

OVERVIEW 

 
The indicator contains two main configuration menus:  

 

The Setup (“F”) menu, which configures the indicator to your weigh platform 

 

The User (“A”) menu, which configures the serial communication port and enables some user 
options 

 

The Setup and User menus consist of several menu selections, each with its own sub-menu of selections 
or programming procedures. To configure the indicator you must first enter the appropriate menu mode. 
Once  there,  four  of  the  front  panel  keys  become  directional  navigators  to  move  around  in  the  menus, 
and one key is used to save or SET the selections. 

 

Summary of Contents for TI-500E SS

Page 1: ...i TI 500E SS Digital Weight Indicator Setup Operation Manual Revision 1 1 July 9 2012...

Page 2: ...ATION F17 11 VIEW CALIBRATION VALUES F18 12 KEY IN ZERO CALIBRATION VALUE F19 13 KEY IN SPAN CALIBRATION VALUE F20 13 OPERATION 14 DISPLAY 14 KEYBOARD 15 FUNCTION KEYS 15 GENERAL SCALE OPERATION 15 WE...

Page 3: ...stallation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications...

Page 4: ...ossible In extreme cases it may be necessary to install surge suppressors line conditioners or even UPS Uninterruptible Power Supplies systems not included Environment Avoid installing the indicator i...

Page 5: ...on the main PCBA Do not connect the load cell cable shield to the green load cell input terminal The TI 500E SS indicator contains a connection terminal on the main board assembly for connection to t...

Page 6: ...cord attached to the indicator The internal power supply is rated for 100 240 VAC and 50 60 Hz Simply plug the unit into a standard wall outlet CONFIGURATION OVERVIEW The indicator contains two main...

Page 7: ...onsist of a top level heading and a secondary level The top level contains the code e g F1 for the parameter to be configured The secondary level contains the selection list or allows access to a prog...

Page 8: ...ty within which the scale may be zeroed Note that the indicator must be in standstill to zero the scale 100 NA 1 9 2 20 F5 Motion Band Sets the level at which motion is detected If motion is not detec...

Page 9: ...n value in case of memory loss in the field Scrolling down with the ZERO key one level begins the procedure Press ZERO key to begin sequence F21 Factory Reset North America This sub menu will reset al...

Page 10: ...ne 4 20 mA output F23 1 While in the Setup Menu mode scroll to F 23 then scroll down once using the ZERO key to enter Fine tune menu The indicator outputs 4 mA and displays a number 2 While monitoring...

Page 11: ...t segments decimal points and LCD annunciators in a test sequence Press ZERO key to begin sequence A6 Output String Selects fixed output string for serial port Standard Chipset only Refer to Appendix...

Page 12: ...he display will show SET momentarily and then revert back up to A8 LF Line Feeds Number Entry A9 1 While in the User Menu mode scroll to A 9 and then scroll down once using the ZERO key to enter the L...

Page 13: ...troubleshoot setup problems 2 After making sure that there are no test weights on the platform press the ZERO key again to zero out the displayed value 3 Press the NET GROSS key to save the zero point...

Page 14: ...causes include incorrect load cell wiring a mechanical obstruction or a faulty load cell VIEW CALIBRATION VALUES F18 Note The values displayed in this procedure are valid only after a successful cali...

Page 15: ...id span calibration values obtained from a successful F17 calibration procedure must be used 1 While in the Setup mode scroll to F 20 and then scroll down once using the ZERO key The indicator will pr...

Page 16: ...ight is active whenever the displayed weight is within a pre programmed band from true zero NET Indicates that the indicator is displaying net weight GROSS Indicates that the indicator is displaying g...

Page 17: ...in motion 3 The scale is not in overload see Appendix D for error codes Print This key is used to send weight information out to the serial port provided the following conditions are met 1 The scale...

Page 18: ...u continue to push the UNITS key the indicator will resort back to weighing mode and you must start again from Step 2 3 Place the sample items on the platform all at once and allow the weight indicati...

Page 19: ...usly and at the same transmission baud rate configured in A1 LEGAL FOR TRADE SEALING Indicators can be sealed for commercial Legal for Trade applications as follows 1 Power off the indicator 2 On the...

Page 20: ...hipset 1024 x 8 internal to C Enhanced Chipset EEPROM 256 x 8 external to C Digital Filtering Software selectable SERIAL COMMUNICATIONS Serial Port Full Duplex selectable Baud rate 8 data bits no pari...

Page 21: ...tor to zero the scale The indicator will not respond if the scale is in motion positive overload or negative overload The indicator will also not respond if it is not in gross mode or within the zero...

Page 22: ...ondec Demand String String Format 1 is designed for two way communication STX POL xxxxx xx LB KG GR NT CR LF Polarity SP Positive Negative Weight Data Units LB pound KG kilogram Gross Net GR Gross NT...

Page 23: ...Keyed in weight value is less than 1 of full scale capacity Use a larger test weight or check keyed in value Err 2 Span Calibration Mode F17 There is not enough load cell signal to produce the intern...