CHAPTER 6: SETUP OPERATION
6.1 SETUP OPERATION OVERVIEW
The system setup consists of three setup areas: Setup Menu, Calibration Menu and Analog to
Digital Conversion Diagnostic Menu.
6.2 SETUP
MENU
The Setup Menu is used to configure the Origination ZIP code, Remote Scale, Filter and Buzzer
etup.
s
1. Press and hold the ZERO key while turning on the scale. When the message “ZIP Code OK?”
appears on the screen, you are in Setup Menu Mode and you may release the ZERO key.
6.2.1 ORIGINATION ZIP CODE
This function is used to program your origination ZIP code.
1. Enter the Setup Menu Mode as described above. The scale prompts ” ZIP Code OK?”
and displays the current ZIP code.
2. Press the ENTER (Yes) key to accept the displayed ZIP code or press the REMOTE
SCALE (No) key to change the ZIP code.
Note:
If
“Yes”
is selected, the scale proceeds to the Remote Scale Setup. If
“No”
is
selected, the message “Input ZIP = ???” appears in the display.
3. Using the numeric keys, enter the first three digits of your ZIP code. Once all three digits
are entered, the scale prompts ” ZIP Code OK?” and displays the current ZIP code.
Repeat Step 2.
6.2.2 REMOTE SCALE ENABLE/DISABLE
This function is used to enable or disable the Remote Scale interface. Note: In order to work
properly, the Transcell SP, SPL or PC Series remote scale must have its serial port
configuration set to the factory default settings. Consult the remote scale’s operation manual
for more information.
1. Enter the Setup Menu Mode as described above and press the ENTER key until the
scale prompts ” Remote Scale OK?” and displays the current status (ON/OFF) of the
remote scale.
2. Press the ENTER (Yes) key to accept the displayed remote scale status or press the
REMOTE SCALE (No) key to change the remote scale status.
Note:
If
“Yes”
is selected, the scale proceeds to Filter Setting. If
“No”
is selected, the
status toggles between ON and OFF.
Page 6-1