6.3 CALIBRATION
MENU
Your SPS scale ships from the factory fully calibrated. There is no need to calibrate the scale
unless you feel that it has become inaccurate. It is a good idea to check the calibration of your
scale from time to time with a precision test weight. Transcell recommends that you perform a new
calibration on your digital scale at least once a year.
The SPS-30 scale may be calibrated using precision test weights of 4 lb and 30 lb. Please note that
if the unit is sealed for use in commercial applications, this calibration procedure is inhibited. See
Chapter 7 for more information.
1. Enter the Calibration Mode by pressing and holding the Library/Book (1) key while turning on
the scale. When the message “Calibration? Y/N” appears on the screen, you are in Calibration
Mode and you may release the Library/Book (1) key.
2. Press the ENTER (Yes) key enter the calibration mode. The scale displays the message “C 0
PR ZERO”.
3. Remove all items from the scale’s platter and then press the ZERO key to zero the scale. The
scale displays the message “C 0 PR ENTER”.
4. Press the ENTER key to save the zero point value. The scale displays the message “C 4 “.
5. Place the 4 lb test weight(s) onto the scale’s platter. When the weight stabilizes, the scale
displays the message “C 4 PR ENTER”.
6. Press the ENTER (Yes) key to save the 4 lb calibration value. The scale displays the message
“C 30 “.
7. Place the 30 lb test weight(s) onto the scale’s platter. When the weight stabilizes, the scale
displays the message “ PR ENTER”.
8. Press the ENTER (Yes) key to save the 30 lb calibration value. The scale then displays the
message “Calibration End”.
9. Turn OFF the scale then turn back ON to resume normal operation.